ColorBox User Guide

Lighting Analysis Dashboards

A dashboard brings together the charts and measurements needed to answer a particular lighting question. Choose Default to understand a new reading, then switch to a task-specific dashboard to see the metrics that matter for camera work, print, human-centric lighting, or a biological application.

In Lighting Analysis → Measurements, select a reading and open the Dashboard selector near its title. Choices depend on the measurement type and the features available to you. Switching dashboards changes what you see and may select a different default Quality profile. The captured spectrum stays the same. When a metric shows Unavailable or Partial, check the input quantity and wavelength coverage shown with the result before using it in a comparison.

Dashboard selector showing general, specialist, and comparison views

See Working with Charts and Metrics for chart controls, spectrum overlays, and metric details.

Default

Use Default for the first review of a spectral reading. It puts white point, output, spectral shape, and color rendition together so you can see where to investigate further.

  • Colorimetry shows correlated color temperature (CCT), Duv, and chromaticity. CCT describes the warm-to-cool appearance of a near-white source. Duv helps explain a tint difference between sources with similar CCT. Use the chromaticity plot to see their position relative to reference loci.
  • Photometry and Spectrum shows the wavelength distribution alongside available output measures. For example, check whether a change in illuminance came with a change in the balance of the blue, green, and red regions.
  • CRI and Fidelity and TM-30 help you inspect how a source renders a range of colors. Use the sample results and hue graphics to see where a summary value hides a particular color shift.

Default dashboard with colorimetry, spectrum, photometry, CRI, and TM-30

Cinema and Camera

Use Camera or Cinema & Camera, depending on the version, to assess sources for film, television, or photography. The dashboard combines exposure-related measurements such as illuminance and foot-candles with CCT, camera-oriented rendition metrics, and spectral similarity.

The SSI display compares the measured spectrum with reference spectra. Use the reference selectors when you need to compare a fixture against a specific daylight or tungsten reference or against another supported light measurement. The spectral plot helps you locate the bands that differ. Equal CCT or illuminance can still leave meaningful differences in the spectrum, which is why these views are useful when matching fixtures for the same scene.

Camera dashboard showing illuminance, color temperature, SSI references, and rendition charts

Lighting Review

Use Lighting Review to assess a fixture, installation, or location in a single view. It brings together CCT, Duv, illuminance, CRI Ra, the spectrum, and chromaticity. The dashboard also shows luminous efficacy of radiation (LER), which expresses how effective the measured radiation is at producing photopic light.

This view helps you explain differences that a single illuminance or CCT value misses. For example, if two samples read similarly on the meter but look different in the room, check Duv and chromaticity for tint, inspect the spectrum for differences in output across wavelengths, then look at rendition for effects on surface colors.

Lighting Review dashboard showing the spectrum, chromaticity, and key fixture metrics

Human-Centric

Use Human-Centric to see how a measured light spectrum relates to photopic vision and the eye's different photoreceptor sensitivities. It places melanopic equivalent daylight illuminance (EDI) beside photopic illuminance and melanopic daylight efficacy ratio (DER). EDI expresses the measured melanopic stimulus in daylight-equivalent lux. DER helps you compare the melanopic contribution per unit of photopic illuminance.

The spectrum and receptor summaries help explain why equal-lux readings can have different melanopic values. Compare readings taken at a consistent position and orientation, such as the same eye-level location under two lighting settings. Where projection controls are available, use them to examine an assumed duration for the measured light.

Human-Centric dashboard with photopic and α-opic results and spectral sensitivity overlays

Use Print Analysis to evaluate a viewing light against a print-viewing reference and preview which colors may change appearance when the light changes. The top cards summarize proximity to D50 in chromaticity, spectral similarity to D50, and near-UV evidence where the instrument covers the required wavelengths. This matters because lights with a similar white point can render colored patches differently.

In the Metamerism preview, set the illuminants to compare, then choose a Patch Set:

  • Synthetic 24 provides a compact set for an initial visual comparison.
  • ColorChecker Classic shows a familiar chart layout for reviewing neutrals, skin-like colors, and saturated patches.
  • ColorChecker Digital SG (Legacy) uses a larger chart so you can inspect more colors.
  • Import loads your own spectral reflectance patches from a CGATS file. After import, choose Custom to see the results for those patches.

Use Visualization to change how the two appearances are presented, such as Side by Side. Turn on Patch ΔE to see the calculated differences, or Patch Numbers to identify particular patches. Use Chart Layout to locate a patch in a familiar grid or order by ΔE to bring the largest differences into view. These previews model appearance from the chosen illuminants and patch reflectance, so read the numerical differences alongside the on-screen colors.

Print Analysis comparing predicted patch appearance with a compact chart

Patch Set menu showing ColorChecker Digital SG and the resulting larger patch grid

Aquarium

Use Aquarium to review spectral and photon delivery for aquatic lighting. Start with Default – physical measurements to inspect the measured spectrum and band totals. Then choose a species profile to bring in the wavelength bands or evidence available for that species. The full-range spectrum helps show which measurements support the results.

Use Profile to browse the catalog, then open Evidence and validation to see the basis and limitations of an entry. Some profiles have a sensitivity curve, while others have more limited evidence. If you need to inspect your own wavelength windows, choose Custom – three analysis bands and set each band's center and offsets. Use Apply photoperiod to project the measured delivery over the selected lighting schedule.

Aquarium dashboard with full-range spectrum, photon metrics, and projection controls

Aquarium Profile menu showing selectable species and evidence groups

Greenhouse

Use Greenhouse to inspect how much light reaches the crop and how photons are distributed across spectral bands. Review photosynthetic photon flux density (PPFD) at the measurement position and use the photoperiod to interpret daily light integral (DLI). The spectrum shows how blue, green, red, far-red, and other measured bands contribute.

Choose a Crop, such as sweet basil, when the analysis should use the available crop-specific context. The dashboard separates physical measurements from outputs about photosynthesis, growth, or leaf chemistry. Use Show evidence where available to understand the source and limits of the selected model before comparing two lighting recipes.

Greenhouse dashboard with a selected crop, photosynthesis view, and spectral bands

Comparison

Use Comparison when the difference between readings matters more than any single reading. Start by selecting a primary reading in the Measurements sidebar, then use the dotted circles to add the other readings to the comparison. A checkmark indicates a reading is included. Choose a reference in Comparison Setup to establish the baseline for the difference values.

Measurements sidebar with a primary reading and two additional readings checked for comparison

The Key Difference Matrix summarizes changes in quantities such as CCT, Duv, output, and rendition. Use the spectral and chromaticity sections to investigate what produced the differences. Compare at the measured level when absolute output matters, or normalize the spectra when you want to focus on shape. Use the same measurement geometry and quantity when comparing absolute output.

Comparison dashboard with multiple measurements visible in the sidebar and the difference matrix

Overlay spectra to locate the bands where fixtures diverge. For example, a match at the blue peak can coexist with a difference across the broader visible spectrum. If curves are difficult to distinguish, click a measurement's color swatch to assign an easier-to-recognize chart color.

Overlaid spectra showing differences across the visible range

Measurement color picker for identifying curves in comparison charts

The CRI comparison shows individual test-color results. Inspect it when two sources have a similar overall CRI but you want to know whether one renders particular colors differently.

Comparison of individual CRI values and color fidelity across samples

Flicker

Select a flicker reading to open the Flicker view. Use the time trace to see how output varies and the frequency views to identify dominant modulation or reported frequency components. The available metrics depend on what the instrument captured and reported.

For camera work, use the exposure controls to examine a frame rate and exposure combination against the available flicker evidence. Compare the predicted exposure modulation with the measured trace and frequency information. This is useful when investigating why a fixture that looks steady to the eye may still cause variation in recorded frames.

Flicker dashboard with camera exposure controls, frequency metrics, and a time trace

Other dashboard choices

Red & NIR

Use Red & NIR when you need to see longer-wavelength output that visible light metrics leave out. Look at the spectrum and band totals to understand where the source delivers radiant power. Check the measurement's wavelength range before comparing two red or near-infrared sources. A visible-range capture can leave part of the output unmeasured.

Poultry

Use Poultry to review spectral and photon balance at a measurement point relevant to the birds, such as bird height. Record the position when comparing areas of an enclosure. Avian receptor measures require a supported profile, and photoperiod or darkness analysis requires a time-based record. A single spot reading describes the site and moment captured.

Manufacturer / R&D

Use Manufacturer / R&D when investigating why a prototype or production sample differs from a target. Look at the spectral diagnostics and reference matching to identify which regions changed before deciding which fixture setting to investigate.

Research / Scientific

Use Research / Scientific when you need a closer look at spectral diagnostics and color fidelity. Keep the observer, input quantities, and wavelength coverage consistent when comparing readings, and use the spectral data to follow up on differences in summary metrics.

Sun Study and Photobiological Hazards

These views open with their dedicated workflows. Sun Study relates a sequence of readings to the time and position of the sun. Photobiological Hazards brings together the available spectral evidence and diagnostics for the selected assessment tasks. Check measurement coverage and missing evidence when interpreting either view.

For spectral reflectance readings, use the separate Reflectance and Material Appearance chapter.